ph-tester-dat

apps

OP-amp-dat for ph-tester-dat

What kind of op-amp IS required?

You need an electrometer-grade or FET-input op-amp.

Suitable op-amps for pH measurement

Op-amp Input bias current
LMP7721-dat 3 fA
LMC6001-dat 25 fA
TL072-dat / TL082-dat ~65 pA (borderline)
OPA129-dat 75 fA
AD8605-dat 1 pA

Rule of thumb:
Input bias current should be < 1/1000 of electrode current


Typical pH front-end architecture

pH electrode
│
│ (very high impedance)
▼
[ FET-input buffer ]
│
├─ Gain & offset
│
├─ Temperature compensation
│
▼
ADC / MCU
  • First stage must be unity-gain buffer
  • Guard rings + clean PCB required
  • Shielded cable mandatory

Ibias

  • Ibias ≤ 1 pA → good pH front end
  • Ibias 1–10 pA → acceptable with care
  • Ibias 10–50 pA → hobby / lab demo only
  • Ibias >50 pA → no